A fundamental question in laboratory medicine is: “How good does my
analytical method need to be?” The answer depends on the intended
clinical use. A method that is acceptable for population screening may
be inadequate for monitoring individual patients. This vignette
introduces the biological variation model for setting analytical
performance specifications, implemented in the valytics
package through three functions: - ate_from_bv(): Calculate
specifications from biological variation data -
sigma_metric(): Quantify performance using the Six Sigma
metric - ate_assessment(): Evaluate observed performance
against specifications
Every measurand (analyte) exhibits natural variation even in healthy individuals. This variation has two components: - Within-subject variation (CVI): Day-to-day fluctuation within an individual - Between-subject variation (CVG): Differences between individuals in a population The biological variation model, developed by Fraser, Petersen, and colleagues, uses these inherent variations to derive meaningful analytical performance goals. The logic is straightforward: analytical error should be small enough that it does not significantly increase the total variation observed in test results.
At the desirable performance level, the formulas are:
Allowable Imprecision: \[CV_A \leq 0.50 \times CV_I\]
Allowable Bias:
\[Bias \leq 0.25 \times \sqrt{CV_I^2 +
CV_G^2}\]
Total Allowable Error:
\[TEa \leq k \times CV_A + Bias\]
Where k is a coverage factor (typically 1.65 for ~95% of
results).
Three performance tiers are defined, each with different multipliers:
| Level | Imprecision | Bias | Stringency |
|---|---|---|---|
| Optimal | 0.25 × CVI | 0.125 × √(CVI² + CVG²) | Most stringent |
| Desirable | 0.50 × CVI | 0.25 × √(CVI² + CVG²) | Standard target |
| Minimum | 0.75 × CVI | 0.375 × √(CVI² + CVG²) | Least stringent |
The ate_from_bv() function calculates all three
specifications from biological variation data:
# Example: Glucose
# CV_I = 5.6%, CV_G = 7.5% (illustrative values)
ate_glucose <- ate_from_bv(cvi = 5.6, cvg = 7.5)
ate_glucose
#>
#> Analytical Performance Specifications from Biological Variation
#> ------------------------------------------------------------
#>
#> Input:
#> Within-subject CV (CV_I): 5.60%
#> Between-subject CV (CV_G): 7.50%
#> Performance level: desirable
#> Coverage factor (k): 1.65
#>
#> Specifications:
#> Allowable imprecision (CV_A): 2.80%
#> Allowable bias: 2.34%
#> Total allowable error (TEa): 6.96%The summary() method shows all three performance
tiers:
summary(ate_glucose)
#>
#> Analytical Performance Specifications - Detailed Summary
#> ============================================================
#>
#> Biological Variation Data:
#> ------------------------------------------------------------
#> Within-subject CV (CV_I): 5.60%
#> Between-subject CV (CV_G): 7.50%
#> Total BV [sqrt(CV_I^2 + CV_G^2)]: 9.36%
#>
#> Settings:
#> ------------------------------------------------------------
#> Selected performance level: desirable
#> Coverage factor (k): 1.65
#>
#> Formulas (Fraser & Petersen 1993):
#> ------------------------------------------------------------
#> CV_A = 0.50 x CV_I
#> Bias = 0.250 x sqrt(CV_I^2 + CV_G^2)
#> TEa = k x CV_A + Bias
#>
#> Specifications (desirable level):
#> ------------------------------------------------------------
#> Allowable imprecision (CV_A): 2.80%
#> Allowable bias: 2.34%
#> Total allowable error (TEa): 6.96%
#>
#> Comparison Across Performance Levels:
#> ------------------------------------------------------------
#> Level CV_A Bias TEa
#> ----- ---- ---- ---
#> optimal 1.40% 1.17% 3.48%
#> desirable * 2.80% 2.34% 6.96%
#> minimum 4.20% 3.51% 10.44%
#>
#> * Selected level
#>
#> Data Source:
#> ------------------------------------------------------------
#> Biological variation values should be obtained from the EFLM
#> Biological Variation Database: https://biologicalvariation.eu/You can calculate specifications for any tier:
If only within-subject variation is available, imprecision goals can still be calculated:
ate_cv_only <- ate_from_bv(cvi = 5.6)
ate_cv_only
#>
#> Analytical Performance Specifications from Biological Variation
#> ------------------------------------------------------------
#>
#> Input:
#> Within-subject CV (CV_I): 5.60%
#> Between-subject CV (CV_G): not provided
#> Performance level: desirable
#> Coverage factor (k): 1.65
#>
#> Specifications:
#> Allowable imprecision (CV_A): 2.80%
#> Allowable bias: requires CV_G
#> Total allowable error (TEa): requires CV_GThe sigma metric provides a standardized way to express method quality. It answers: “How many standard deviations of analytical error fit between my observed performance and the allowable limit?” \[\sigma = \frac{TEa - |Bias|}{CV}\]
Higher sigma values indicate better performance:
| Sigma | Category | Defects per Million |
|---|---|---|
| ≥ 6 | World Class | ~3.4 |
| ≥ 5 | Excellent | ~230 |
| ≥ 4 | Good | ~6,200 |
| ≥ 3 | Marginal | ~66,800 |
| ≥ 2 | Poor | ~308,500 |
| < 2 | Unacceptable | > 690,000 |
# Assume observed: bias = 1.5%, CV = 2.5%
# Using TEa from biological variation
sm <- sigma_metric(
bias = 1.5,
cv = 2.5,
tea = ate_glucose$specifications$tea
)
sm
#>
#> Six Sigma Metric
#> ----------------------------------------
#>
#> Input:
#> Observed bias: 1.50%
#> Observed CV: 2.50%
#> Total allowable error (TEa): 6.96%
#>
#> Result:
#> Sigma: 2.18
#> Performance: Poor
#> Defect rate: ~308,500 per millionsummary(sm)
#>
#> Six Sigma Metric - Detailed Summary
#> ==================================================
#>
#> Formula:
#> --------------------------------------------------
#> Sigma = (TEa - |Bias|) / CV
#> Sigma = (6.96 - |1.50|) / 2.50
#> Sigma = (6.96 - 1.50) / 2.50
#> Sigma = 5.46 / 2.50
#> Sigma = 2.18
#>
#> Result:
#> --------------------------------------------------
#> Sigma metric: 2.18
#> Performance category: Poor
#> Expected defect rate: ~308,500 per million
#>
#> Sigma Scale Reference:
#> --------------------------------------------------
#> Sigma Category Defects/Million
#> ------ ------------- ---------------
#> >= 6 World Class 3.4
#> >= 5 Excellent 230
#> >= 4 Good 6,210
#> >= 3 Marginal 66,800
#> >= 2 Poor 308,500 *
#> < 2 Unacceptable > 690,000
#>
#> * Current performance level
#>
#> Note: Defect rates assume 1.5 sigma long-term shift.In clinical laboratories:
The ate_assessment() function evaluates observed
performance against specifications:
assess <- ate_assessment(
bias = 1.5,
cv = 2.5,
tea = ate_glucose$specifications$tea
)
assess
#>
#> Analytical Performance Assessment
#> --------------------------------------------------
#>
#> >>> METHOD ACCEPTABLE <<<
#>
#> Performance Summary:
#> Parameter Observed Allowable Status
#> -------------------- ---------- ---------- ----------
#> Bias 1.50% --- ---
#> CV (Imprecision) 2.50% --- ---
#> Total Error 5.62% 6.96% PASS
#>
#> Sigma Metric: 2.18 (Poor)When you have specifications for all components:
assess_full <- ate_assessment(
bias = 1.5,
cv = 2.5,
tea = ate_glucose$specifications$tea,
allowable_bias = ate_glucose$specifications$allowable_bias,
allowable_cv = ate_glucose$specifications$allowable_cv
)
summary(assess_full)
#>
#> Analytical Performance Assessment - Detailed Summary
#> ============================================================
#>
#> Overall Result: METHOD ACCEPTABLE
#>
#> Observed Performance:
#> ------------------------------------------------------------
#> Bias: 1.50%
#> CV (Imprecision): 2.50%
#> Total Error (k=1.65): 5.62%
#> [TE = 1.65 x 2.50 + |1.50| = 5.62]
#>
#> Allowable Specifications:
#> ------------------------------------------------------------
#> Allowable Bias: 2.34%
#> Allowable CV: 2.80%
#> Total Allowable Error (TEa): 6.96%
#>
#> Component Assessment:
#> ------------------------------------------------------------
#> Bias: PASS (margin: +0.84%)
#> |1.50| <= 2.34
#> CV: PASS (margin: +0.30%)
#> 2.50 <= 2.80
#> Total Error: PASS (margin: +1.34%)
#> 5.62 <= 6.96
#>
#> Sigma Metric:
#> ------------------------------------------------------------
#> Sigma = (TEa - |Bias|) / CV
#> Sigma = (6.96 - 1.50) / 2.50 = 2.18
#> Category: Poor
#>
#> Sigma Scale:
#> >= 6: World Class | >= 5: Excellent | >= 4: Good
#> >= 3: Marginal | >= 2: Poor | < 2: Unacceptable
#>
#> Interpretation:
#> ------------------------------------------------------------
#> Method technically passes but sigma < 3 indicates high risk.
#> Strongly recommend method improvement or enhanced QC.# A method with poor performance
assess_poor <- ate_assessment(
bias = 4.0,
cv = 5.0,
tea = ate_glucose$specifications$tea,
allowable_bias = ate_glucose$specifications$allowable_bias,
allowable_cv = ate_glucose$specifications$allowable_cv
)
summary(assess_poor)
#>
#> Analytical Performance Assessment - Detailed Summary
#> ============================================================
#>
#> Overall Result: METHOD NOT ACCEPTABLE
#>
#> Observed Performance:
#> ------------------------------------------------------------
#> Bias: 4.00%
#> CV (Imprecision): 5.00%
#> Total Error (k=1.65): 12.25%
#> [TE = 1.65 x 5.00 + |4.00| = 12.25]
#>
#> Allowable Specifications:
#> ------------------------------------------------------------
#> Allowable Bias: 2.34%
#> Allowable CV: 2.80%
#> Total Allowable Error (TEa): 6.96%
#>
#> Component Assessment:
#> ------------------------------------------------------------
#> Bias: FAIL (margin: -1.66%)
#> |4.00| > 2.34
#> CV: FAIL (margin: -2.20%)
#> 5.00 > 2.80
#> Total Error: FAIL (margin: -5.29%)
#> 12.25 > 6.96
#>
#> Sigma Metric:
#> ------------------------------------------------------------
#> Sigma = (TEa - |Bias|) / CV
#> Sigma = (6.96 - 4.00) / 5.00 = 0.59
#> Category: Unacceptable
#>
#> Sigma Scale:
#> >= 6: World Class | >= 5: Excellent | >= 4: Good
#> >= 3: Marginal | >= 2: Poor | < 2: Unacceptable
#>
#> Interpretation:
#> ------------------------------------------------------------
#> Method does not meet specifications.
#> - Bias exceeds allowable limit: consider recalibration.
#> - Imprecision exceeds allowable limit: investigate sources.
#> - Total error exceeds TEa: method requires improvement.Here is a typical workflow for evaluating a new glucose method:
# Step 1: Define quality goals from biological variation
specs <- ate_from_bv(cvi = 5.6, cvg = 7.5, level = "desirable")
cat("Quality Specifications:\n")
#> Quality Specifications:
cat(sprintf(" Allowable CV: %.2f%%\n", specs$specifications$allowable_cv))
#> Allowable CV: 2.80%
cat(sprintf(" Allowable Bias: %.2f%%\n", specs$specifications$allowable_bias))
#> Allowable Bias: 2.34%
cat(sprintf(" TEa: %.2f%%\n\n", specs$specifications$tea))
#> TEa: 6.96%
# Step 2: Assume we measured method performance
# (In practice, from validation studies)
observed_bias <- 1.8
observed_cv <- 2.2
# Step 3: Calculate sigma metric
sm <- sigma_metric(observed_bias, observed_cv, specs$specifications$tea)
cat(sprintf("Sigma Metric: %.2f (%s)\n\n", sm$sigma, sm$interpretation$category))
#> Sigma Metric: 2.35 (Poor)
# Step 4: Full assessment
assessment <- ate_assessment(
bias = observed_bias,
cv = observed_cv,
tea = specs$specifications$tea,
allowable_bias = specs$specifications$allowable_bias,
allowable_cv = specs$specifications$allowable_cv
)
# Step 5: Decision
if (assessment$assessment$overall) {
cat("DECISION: Method acceptable for clinical use\n")
} else {
cat("DECISION: Method requires improvement\n")
}
#> DECISION: Method acceptable for clinical useThe quality of your specifications depends on reliable biological variation estimates.
The EFLM Biological Variation Database is the current authoritative source:
While the biological variation model is widely used, it is not the only approach to setting quality specifications. Other models include:
The ate_assessment() and sigma_metric()
functions work with TEa values from any source—simply provide your
specification directly rather than calculating from biological
variation.
# Using a CLIA-based TEa for glucose (example: ±6 mg/dL or ±10%)
# For a sample at 100 mg/dL, 10% = 10 mg/dL
sm_clia <- sigma_metric(bias = 2, cv = 3, tea = 10)
sm_clia
#>
#> Six Sigma Metric
#> ----------------------------------------
#>
#> Input:
#> Observed bias: 2.00%
#> Observed CV: 3.00%
#> Total allowable error (TEa): 10.00%
#>
#> Result:
#> Sigma: 2.67
#> Performance: Poor
#> Defect rate: ~308,500 per millionThe biological variation model provides a scientifically grounded approach to setting analytical quality specifications:
ate_from_bv() translates biological
variation into actionable specificationssigma_metric() provides a universal
quality scale for comparing methodsate_assessment() gives a clear
pass/fail evaluationThese tools help laboratories make informed decisions about method acceptability while recognizing that the final decision depends on clinical context and regulatory requirements.
Fraser CG, Petersen PH (1993). Desirable standards for laboratory tests if they are to fulfill medical needs. Clinical Chemistry, 39(7):1447-1453.
Ricos C, Alvarez V, Cava F, et al. (1999). Current databases on biological variation: pros, cons and progress. Scandinavian Journal of Clinical and Laboratory Investigation, 59(7):491-500.
Aarsand AK, Fernandez-Calle P, Webster C, et al. (2020). The EFLM Biological Variation Database. https://biologicalvariation.eu/
Westgard JO, Westgard SA (2006). The quality of laboratory testing today: an assessment of sigma metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for acceptable performance. American Journal of Clinical Pathology, 125(3):343-354.